ProTime Microcoagulation System

The ProTime Microcoagulation System is designed to safely monitor the clotting activity in blood in patients on warfarin anticoagulant therapy. The test procedure is simple and safe with lab quality performance.
 

  • CLIA-waived
  • Accepts and stores patient and/or operator identification (ID) numbers which can eliminate transcription errors
  • Auto-sends results quickly and easily to printer, computer or LIS
  • Only system offering the flexibility of both on-board and optional external quality controls
  • ISI of thromboplastin reagent = 1.0 (AHA, CAP, and WHO recommended)
  • Off-meter sampling provides safety against disease transmission
  • Available for patient self-testing

 

To learn more about ProTime and Anticoagulation Therapy, please select from the links below:

Anticoagulation Therapy
Performing a Finger-stick with Tenderlett® Plus
Running a PT/INR Test
Changing the Battery
Running Liquid Quality Control (For Professional Use)
Entering PID/OID (For Professional Use)

Back to Product List

Test menu

  • Prothrombin Time Test (PT/INR)

Consumable

ProTime 5 Channel CuvetteProTime 3 Channel Cuvette

ProTime 5 Channel Cuvette ProTime 3 Channel Cuvette
  ISI ~ 1.0   ISI ~ 1.0
  Barcoded   Barcoded
  16 hour open-pouch stability   16 hour open-pouch stability
  Triplicate patient testing   Single patient test
  On-board Level I and Level II control   On-board Level I and Level II control
  Sample size 67 µL   Sample size 27 µL


Optional external quality controls (catalog # DCPRO-N and DCPRO-A) are available for the
ProTime Microcoagulation System.

Specifications

Sample Size 27µL (Pro 3 Cuvette) / 67µL (Pro 5 Cuvette)
CLIA Status Waived
Reportable Range (INR) 0.8 - 9.9
Detection Technology Mechanical clot based technology - optical detection
External LQC (Optional) Yes
Reagent Stability (Out of Pouch) 16 hrs
Shelf-Life (Room Temp) 60 days
Lot Code Entry Automatic (barcode)
Memory 50 results
PID/OID Yes
Interface 2 button
Warranty 3 year