IRMA TruPoint Blood Analysis System

IRMA TruPoint offers bedside testing that is accurate, reliable and self-contained.

 

Accurate

  • Precise results within two minutes
  • Lab quality accuracy
  • Comprehensive menu
  • Custom correlation

Reliable

  • Single-use cartridge for maintenance free convenience
  • Self-contained calibrant in each cartridge
  • Real-time, automatic electronic quality control

Self-Contained

  • All-in-one design is lightweight and battery-operated
  • On-board thermal printer documents results and test notes
  • Interactive touch screen

 

Watch the IRMA TruPoint Training Video here.

 

Back to Product List

Test menu

Measured Values:

  • BG Cartridge: pH, pCO2, pO2
  • H3 Cartridge: Na+, K+, iCa, Hct
  • CC Cartridge: pH, pCO2, pO2, Na+, K+, iCa, Hct
  • H4 Cartridge: Na+, K+, Cl-, BUN, Hct
  • GL Cartridge: Na+, K+, Cl-, Glucose
  • CR Cartridge: Creatinine
  • LA Cartridge: Lactate


Calculated Values:

  • BG Cartridge: HCO3-, TCO2, BEb, BEecf, O2SAT
  • H3 Cartridge: tHb
  • CC Cartridge: HCO3-, TCO2, BEb, BEecf, O2SAT, tHb, iCa(N)
  • H4 Cartridge: tHb
  • CR Cartridge: MDRD-GFR

Consumable

  • Room Temperature Storage (Creatinine cartridge requires refrigeration)
  • Single-use cartridge
  • Calibration with every test
  • No external reagents or gases required
  • Closed sample introduction system
  • Easy to use

 

Cartridge Test Menu - Measured Values:

  CC BG H3 H4 GL CR LA
pH  ♦  ♦          
pCO2  ♦  ♦          
pO2  ♦  ♦          
Na+  ♦    ♦  ♦  ♦    
K+  ♦    ♦  ♦  ♦    
Cl-        ♦  ♦    
iCa  ♦    ♦        
Glu          ♦    
BUN        ♦      
Hct  ♦    ♦  ♦      
Creatinine            ♦  
Lactate              ♦

 

Cartridge Test Menu - Calculated Values:

  CC BG H3 H4 GL CR
HCO3-  ♦  ♦        
TCO2  ♦  ♦        
BEecf  ♦  ♦        
BEb  ♦  ♦        
O2SAT  ♦  ♦        
iCa(N)  ♦          
tHb  ♦    ♦  ♦    

 

Specifications

Reportable Ranges Measured:

pH 6.000-8.000 pH units
pCO2 4.0-200.0 mmHg (0.53-26.66 kPa)
pO2 20.0-700.0 mmHg (2.67-93.33 kPa)
Na+ 80.0-200.0 (mM, mEq/L)
K+ 1.00-20.00 (mM, mEq/L)
Cl- 30.0-150.0 (mM, mEq/L)
iCa 0.20-5.00 mM
(0.40-10.00 mEq/L; 0.80-20.00 mg/dL)
BUN 3-150 mg/dL (1.1-53.5 mM)
Glucose 20-500 mg/dL (1.1-27.8 mM)
Hct 10.0-80.0% (.100-.800 SI)
Creatinine 0.2-12 mg/dL (18-1061 µmol/L)
Lactate 0.30-12 mmol/L (2.7-108 mg/dL)

 

General Specifications:

Minimum Blood Sample Size 0.2mL from a syringe
0.125mL from the IRMA TruPoint
Capillary Collection Device
Data Storage 200 patient test records
150 liquid QC records
100 electronic QC / Temperature test results
Cartridge Storage Temperature 15-30°C / 59-86°F
Creatinine Cartridge
Storage Temperature
2-8°C / 35.6-46.4°F

 

 

 

Other

Optional Accessories:

  • Barcode reader
  • idms™ data management software
  • Device Set configuration software

 

Connectivity:

Interface connectivity available through:

  • AegisPOC™
  • RALS®-Plus
  • TELCOR QML®

 

Quality Control:

  • Liquid Quality control performed using RNA® Medical Controls. For detailed information, visit RNA Medical Controls